Product analysis: analysis was able to confirm that the external pulse generator (epg) displayed an error code; a capacitor on the main printed circuit board (pcb) was broken off.Analysis also noted that the upper case was cracked at the boss, the lower case was broken at the boss, and the display wires were pinched with exposed wires.Additionally, the encoder assembly, one knob encoder, the main seal, and one case screw were all contaminated.All found defective parts were replaced and all other identified issues were resolved.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
|