| Model Number |
8000.COM03 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre b.V.).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(6) 2018.Since this incident occurred during surgery.It can lead to hypotony.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
Machine suddenly stopped responding, the screen was frozen, it was aspirating and irrigating although no one was pressing the footswitch, there was no error message and it was impossible to turn off the machine.After rebooting the eva directly from the main switch on the back, the system entered the demo mode every time.We tried to reboot it plenty of times but it kept entering the demo mode.
|
| |
|