| Model Number |
01-0043R |
| Medical Device Problem Codes |
False Positive Result (1227); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to use, the unit would shutdown often and when it scan, it got wrong readings.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device sample was received for evaluation.From the provided information, it is unknown if the device has or has not been used in the treatment or diagnosis of a patient.The returned sample did not meet specification as received.The visual inspection found no notable conditions.The investigation isolated the failure to a software issue.The software was reloaded to address the condition.Corrective actions have been implemented to mitigate this issue.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|