Model Number 840
Medical Device Problem Code Failure To Run On AC/DC (1001)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: the field service engineer evaluated the ventilator and replaced the power supply.The ventilator passed all tests and calibrations; operates within medtronic specifications upon completion of service.The tamura (taiyo yuden) power supply was returned for failure investigation.A visual inspection of the returned part was conducted and no anomalies were observed.The part was attached to the failure investigation test ventilator for analysis.The ventilator was powered up bit failed at the power up stage with error codes.The fault was isolated to the v3 (+15v) and v4 (-15v) rails to the bdu breath delivery unit.The voltage rails were out of specification, reading approximately +6v and +0.6v respectively, when they should have been reading +15v and -15v respectively.All other voltage rails were reading within specification.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that an 840 ventilator generated ventilator inoperable error codes.The ventilator was not in use on a patient at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7664899
Report Number8020893-2018-00302
Device Sequence Number2514876
Product Code CBK
UDI-Device Identifier10884521077188
UDI-Public10884521077188
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840220DRAC-GR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2018
Initial Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-