Device evaluation summary: the field service engineer evaluated the ventilator and replaced the power supply.The ventilator passed all tests and calibrations; operates within medtronic specifications upon completion of service.The tamura (taiyo yuden) power supply was returned for failure investigation.A visual inspection of the returned part was conducted and no anomalies were observed.The part was attached to the failure investigation test ventilator for analysis.The ventilator was powered up bit failed at the power up stage with error codes.The fault was isolated to the v3 (+15v) and v4 (-15v) rails to the bdu breath delivery unit.The voltage rails were out of specification, reading approximately +6v and +0.6v respectively, when they should have been reading +15v and -15v respectively.All other voltage rails were reading within specification.If information is provided in the future, a supplemental report will be issued.
|