TERUMO CORPORATION, ASHITAKA CAPIOX RX OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX*RX15RW30
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D6: implanted date: device was not implanted d7: explanted date: device was not explanted the actual device was returned for evaluation.Visual inspection upon receipt found that the cap had been removed from the blood outlet port and the blood outlet port had been crushed inward from the outside.The cap was not retuned.Reproductive testing was performed.The blood outlet port on an oxygenator sample, with the cap attached to it, was hit on a hard object.As the result, the cap and the blood outlet port got deformed.The configuration of deformation on the blood outlet port was confirmed to be similar to that observed on the actual sample.A review of the device history record and the product release decision control sheet of the reported product code/lot number combination was conducted with no findings.Ifu reference: inspect the device and package carefully.Do not use if the package and/or device is damaged.There is no evidence that this event was related to a device defect or malfunction.Based on the investigation result, it is likely that the actual sample was exposed to some hitting force on the blood outlet port on the oxygenator module, resulting in the reported damage.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.Exemption number e2015022.
 
Event or Problem Description
The user facility reported when the perfusionist was opening the immediate capiox device box for surgery, they saw a squeezed area in the blood outlet area of the oxygenator.The event occurred pre-treatment and the patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX OXYGENATOR
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, shizuoka 418
JA  418
MDR Report Key7666176
Report Number9681834-2018-00124
Device Sequence Number1040399
Product Code DTZ
UDI-Device Identifier04987350771056
UDI-Public04987350771056
Combination Product (Y/N)N
Initial Reporter CountryCE
PMA/510(K) Number
K051997
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberCX*RX15RW30
Device Lot Number170519
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2018
Initial Report FDA Received Date07/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-