• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US RAPIDLOC (PLA/PDS) 2/0 PANACRYL 27° CURVE NEEDLE; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US RAPIDLOC (PLA/PDS) 2/0 PANACRYL 27° CURVE NEEDLE; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 228312
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/31/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement the complaint device is not being returned, therefore is unavailable for a physical evaluation.Furthermore, no lot numbers were supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.Therefore, we cannot determine what caused the user to experience the reported event, we cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).Depuy synthes has been informed that the lot number is not available.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.Associated medwatch: 1221934-2018-51453.
 
Event or Problem Description
It was reported by the sales rep that the sutures on two of the customer's rapidloc 27 degree needle with panacryl broke while the surgeon was pulling tension to push the knots down during a meniscal procedure.This happened one after the other and the surgeon removed both anchors after the sutures broke and replaced them with 12 degree needle anchors.There were no patient consequences.There was a ten minute delay.Both devices were discarded by the customer already.The sales rep was not present, no further details were available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDLOC (PLA/PDS) 2/0 PANACRYL 27° CURVE NEEDLE
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7667113
Report Number1221934-2018-51452
Device Sequence Number19092668
Product Code MAI
UDI-Device Identifier10886705010202
UDI-Public10886705010202
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/01/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number228312
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/01/2013
Initial Report FDA Received Date07/06/2018
Patient Sequence Number1
-
-