• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US PANALOK RC QUICK ANCHOR PLUS, W/DUAL SUTURE PANACRYL; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US PANALOK RC QUICK ANCHOR PLUS, W/DUAL SUTURE PANACRYL; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 212732
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.The batch review 3694720 showed that the devices were conformed to in process and finished goods specifications when released to stock the 29 apr.2013.Expiration date: 2016-04 (according to retained labels).No nc was opened.The complaint device was received and evaluated.Visual inspection revealed that only inserter and suture were received.The anchor itself was not received.There was tissue debris on the inserter and the suture.We cannot confirm this complaint or discern a root cause for the reported failure without the anchor.Typically, from historical investigations, anchor breaks are associated with use of incorrect instruments, levering during insertion or axial misalignment.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The procedure was rcr.Anchor broke when surgeon hit it in before any tying of suture.No adverse effect to patient, other than extra anesthetic time.There was five (5) minutes delay in procedure, just to get a new anchor.Broken items all removed and another anchor inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANALOK RC QUICK ANCHOR PLUS, W/DUAL SUTURE PANACRYL
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7667424
Report Number1221934-2018-51730
Device Sequence Number11025537
Product Code MAI
UDI-Device Identifier10886705002184
UDI-Public10886705002184
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/30/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2016
Device Catalogue Number212732
Device Lot Number3694720
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2013
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/30/2013
Initial Report FDA Received Date07/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-