| Model Number |
APTIO BY SIEMENS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted the siemens customer care center (ccc) to report that a low creatinine (cre) test result was obtained on an aliquot from the primary sample tube on an aptio by siemens system.The customer found water in the aliquoter secondary tube on the aptio input/output module.The customer turned off the aliqoter.A siemens customer service engineer (cse) was dispatched to the site to inspect the system.The cse replaced a pipetter valve and pump.The cause of the low creatinine test result was the pipette valve and pump.The instrument is performing according to specifications.No further evaluation of this instrument is required.
|
| |
|
Event or Problem Description
|
|
A low creatinine (cre) test result was obtained on an aliquot from the primary sample tube on an aptio by siemens system.The initial result was reported to the physician(s).The primary sample was repeated and the repeat test result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the low creatinine test result.
|
| |
|