| Model Number |
48393110 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The cage tail broke during insertion.No response was received from the stryker representative regarding additional information.No adverse consequences or surgical delay were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assessment.Results: the event was confirmed via email correspondence with the sales representative.Visual, dimensional and functional analysis could not be performed as the device was not returned.No lot # was provided, so a manufacturing record review could not be performed.Conclusion: too much impaction force on the inserter may damage the inserter and break the implant wings.Per email correspondence with the sales rep, the implant was impacted using a mallet with a normal force - no extra force was applied.The product was not returned for evaluation, therefore exact root cause cannot be determined conclusively.
|
| |
|
Event or Problem Description
|
|
The cage tail broke during insertion.No response was received from the stryker representative regarding additional information.No adverse consequences or surgical delay were reported.
|
| |
|