No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, the device history records is currently under review.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found the balloon not detached from the catheter shaft with no signs of retraction issue to the balloon or catheter shaft.Therefore, the investigation is unconfirmed for the reported detachment or retraction issue.However, during functional testing a rupture was identified in the proximal cone of the balloon.The definitive root cause for the identified rupture could not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H11: h3, h6 (device code + description: removed 1536, added 2907, device code + description3: removed 2907, eval code & desc - results 1, conclusion1) h11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|