Catalog Number CRU14SA
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one crosser catheter was returned for evaluation.A complete circumferential break was noted in the strain relief, exposing the core wire.Functional testing was not performed due to sample condition.The investigation is therefore confirmed for the identified break and unconfirmed for a crack, as the break was most likely the identified crack.The investigation is inconclusive for the alleged leak, as functional testing could not be performed due to sample condition.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Expiry date - 04/2020.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during the saline flush portion of the out of body test for a recanalization catheter, an alleged leak was noted and a break was found at the distal end of the handle.Reportedly, another recanalization catheter was used to complete the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7668173
Report Number2020394-2018-01147
Device Sequence Number13733334
Product Code PDU
UDI-Device Identifier10801741125567
UDI-Public(01)10801741125567
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCRU14SA
Device Lot NumberGFCQ2862
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2018
Initial Report FDA Received Date07/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-