| Model Number |
125028 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the fiber on tubing lumen side seemed to be stuck in.The facility operate hyperhidrosis surgery with the product more than 10 par month.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample returned for evaluation contained in a plastic bag without its original unit pack.Visual examination shows no sign of any defect.The occlusion test was carried out on the returned unit.The tip of the rod must pass through the y-adaptor for a minimum distance of 1cm.The rod passed freely through the y-adaptor and tubing.This confirms that there was no occlusion present on the unit returned for evaluation.The most probable root cause is the cuff body came in contact with a sharp utensil or object.The most probable root cause is the end user use a scope with a diameter which was not compatible with the diameter of our tubing.All of our product undergo a 100% occlusion test during the manufacturing process.It is at this stage process that occlusions/blockages in the main stem of the tubing are identified and removed from the lot.The reported defect could not be detected on the unit returned for evaluation.The device failed to meet specification as it was received or made available for evaluation.The cause of the observed condition was determined to be cuff body came in contact with a sharp utensil or object.The product relates to the reported complaint event.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|