• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-722NAS
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed button error.The customer¿s blood glucose was 146mg/dl at the time of incident.Customer stated that the insulin pump alarmed button error and had unresponsive buttons.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is not advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7670282
Report Number3004209178-2018-89257
Device Sequence Number16789478
Product Code OYC
UDI-Device Identifier00613994312587
UDI-Public(01)00613994312587
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-722NAS
Device Catalogue NumberMMT-722NAS
Device Lot NumberA3722NASJ
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2018
Initial Report FDA Received Date07/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
Patient Weight256
-
-