| Catalog Number |
242204000 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Today during surgery the surgeon in (b)(6) needed to use this instrument.On attempt to put this together he felt the instrument was weak at the top and did not believe the threads to be fitting together properly.The two parts of the slap hammer that would not fit together are below.In addition due to this problem he was not able to remove a stem in situ, and the surgery was extended for 2 hours while we got another extraction kit from a local hospital.When using the same item from the new kit the stem came straight out.During surgery today when using the slap hammer, on removal of the stem the stem would not come off the end of the slap hammer.They seemed "welded" together.The surgeon was unhappy with this and deemed it to be a fault of the slap hammer.In addition he felt that this item was weak in removing the construct and believed the threads to cross if the slap hammer was used too vigorously, which is the nature of the tool being used.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The investigation could not draw any conclusions regarding the reported event.Based on the inability to determine a root cause, the need for corrective action was not indicated.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|