| Model Number |
419578 |
| Medical Device Problem Code |
Capturing Problem (2891)
|
| Health Effect - Clinical Codes |
Muscle Stimulation (1412); Dyspnea (1816)
|
| Date of Event |
02/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 693565 lead implanted: (b)(6) 2010; 383059 lead implanted: (b)(6) 2010.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the left ventricular (lv) lead displayed increased thresholds which affected the battery life of the cardiac resynch ronization therapy defibrillator (crt-d).The lv lead was inactivated however the patient experienced dyspnea on exertion and the lead was turned back on.It was further reported there was chronic left diaphragmatic stimulation.As the stimulation did not bother the patient, only the generator change was performed due to elective replacement indicator (eri).The lead remains in use and the device was explanted and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|