MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problem Device Difficult to Setup or Prepare (1487)
Patient Problem No Patient Involvement (2645)
Event Date 06/11/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Device manufacturing date is unavailable.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the camera was not working properly.No patient was present.
 
Manufacturer Narrative
Additional information: manufacture date provided.
 
Manufacturer Narrative
Additional information: the logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.Correction: disregard supplemental report 1 as it was intended for mdr 1723170-2018-04847.
 
Manufacturer Narrative
Additional information: a medtronic representative went to the site to test the equipment.Testing revealed that the camera was turning on and off.Replacing the camera resolved the issue.The system then passed the system checkout and was found to be fully functional.The camera has not been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7671284
MDR Text Key113502785
Report Number1723170-2018-03243
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9735665
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse