Catalog Number 9735665 |
Device Problem
Device Difficult to Setup or Prepare (1487)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Device manufacturing date is unavailable.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the camera was not working properly.No patient was present.
|
|
Manufacturer Narrative
|
Additional information: manufacture date provided.
|
|
Manufacturer Narrative
|
Additional information: the logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.Correction: disregard supplemental report 1 as it was intended for mdr 1723170-2018-04847.
|
|
Manufacturer Narrative
|
Additional information: a medtronic representative went to the site to test the equipment.Testing revealed that the camera was turning on and off.Replacing the camera resolved the issue.The system then passed the system checkout and was found to be fully functional.The camera has not been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
|