• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP PROXIMAL OLECRANON PL 2H/RT/73MM-STER; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP PROXIMAL OLECRANON PL 2H/RT/73MM-STER; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.107.002S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Nonunion/Delayed-union Bone Fracture (2369); Impaired Healing (2378)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient¿s height reported as 162 cms.Additional patient identifier reported as (b)(6).Date of event: date of postoperative non-union development is unknown.Additional device product code: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested and is currently pending completion.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent osteotomy of the olecranon on (b)(6) 2018 to fix the humerus.The osteotomy was repaired with an olecranon plate, 2 cortex screws, and 4 variable angle (va) locking screws.On (b)(6) 2018 the patient had a removal of variable angle locking compression plate due to non-union.It was noted that the proximal olecranon did not heal and the piece of bone came out of the screw plate construct, and patient also experienced malunion and decreased range of motion.There was no other medical intervention required.There was no surgical delay reported.The procedure was successfully completed.The patients' proximal olecranon was removed and the patient outcome was stable.This is report 1 of 7 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: device evaluated by mfr, device manufacture date.Device history record (dhr) review: part#: 02.107.002s; lot#: h593743; mfg location: elmira synthes usa; mfg date: 02 april 2018; expiration date: 28 february 2028; noted nonconformances: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.7mm/3.5mm va-lcp proximal olecranon product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM/3.5MM VA-LCP PROXIMAL OLECRANON PL 2H/RT/73MM-STER
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7671411
Report Number2939274-2018-52749
Device Sequence Number14098484
Product Code HRS
UDI-Device Identifier10886982029232
UDI-Public(01)10886982029232
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K120070
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.107.002S
Device Catalogue Number02.107.002S
Device Lot NumberH593743
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/12/2018
Supplement Date Received by Manufacturer07/24/2018
Initial Report FDA Received Date07/09/2018
Supplement Report FDA Received Date07/30/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient Weight78
-
-