BECTON DICKINSON MEDICAL SYSTEMS UNSPECIFIED BD POSIFLUSH¿ NORMAL SALINE SYRINGE; PREFILLED SALINE FLUSH SYRINGE Back to Search Results
Catalog Number UNKNOWN
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Bacterial Infection (1735); Fever (1858); Headache (1880); Unspecified Infection (1930); Nausea (1970); Chills (2191)
Event Date 06/18/2018
Event Type  Injury  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event Description
It was reported that a patient has experienced recurring headaches with nausea, chills, and fever beginning (b)(6) 2017.The patient went to an emergency department and received an mri.The patient associates her symptoms with the use of an unspecified bd posiflush¿ normal saline syringe that was administered at a pain clinic.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the (b)(6) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the (b)(6) identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing (b)(6) for this bacterium.Investigations are ongoing by bd, fda, and cdc.
 
Manufacturer Narrative
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: a.3.Sex: female.B.2.Event attributed to: required intervention.D.8.Device single use?: no.H.3.Device returned to manufacture: no.Investigation summary: as both a lot number and sample were unavailable for this incident, a full investigation consisting of a sample analysis and a device history record review could not be completed.Prior to april 2018, there was no trend for infection or reaction regarding the bd franklin products.Upon the increase of infection and reaction reports beginning in april 2018, a corrective and preventive action plan was initiated to address the issue.A review of all lot sterility testing performed for product released between april 2015 and may 2018 confirmed that no organism growth was identified for any lots released during the shelf life of this product.A direct causation between the reported reaction and the bd franklin product has not been identified.Prior to the report of infection evaluated in mps-18-1248-sa, there was no trend for reaction cases reported for franklin product.This complaint is part of a new trend which began in april 2018.Capa 350041 was initiated to address this issue.As no lot number was provided, a dhr review could not be completed.Samples and/or pictures were not provided for evaluation.A review of all lot sterility testing performed for product released between april 2015 & may 2018 confirmed that no organism growth was identified for any lots released during the shelf life of this product.The root cause analysis of the reported infection cases under capa 350041 has not identified a direct causation between the infections and the bd franklin product.
 
Event Description
"it was reported that a patient has experienced recurring headaches with nausea, chills, and fever beginning october-november 2017.The patient went to an emergency department and received an mri.The patient associates her symptoms with the use of an unspecified bd posiflush¿ normal saline syringe that was administered at a pain clinic.*out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
PREFILLED SALINE FLUSH SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
9630 south 54th street
franklin WI 53132
MDR Report Key7672136
MDR Text Key113511075
Report Number2134319-2018-00083
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-