• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device powered off by itself during a patient event.This issue occurred when taking a blood pressure measurement and a 12 lead ecg reading.This issue is patient related; however there was no adverse patient outcome reported by the customer.Physio-control made multiple attempts to contact the customer in order to obtain additional information about both the patient and the event, but has been unsuccessful.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer¿s device but was unable to duplicate the reported issue.Physio downloaded the electronic records from the customer¿s device and was able to confirm that their device experienced multiple inappropriate losses of power.It was observed that the battery used during the event had a broken locking tab.The device batteries were replaced, and then after other unrelated repairs were completed, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
 
Event or Problem Description
The customer contacted physio-control to report that their device powered off by itself during a patient event.This issue occurred when taking a blood pressure measurement and a 12 lead ecg reading.This issue is patient related; however there was no adverse patient outcome reported by the customer.Physio-control made multiple attempts to contact the customer in order to obtain additional information about both the patient and the event, but has been unsuccessful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7672155
Report Number0003015876-2018-01063
Device Sequence Number2643229
Product Code MKJ
UDI-Device Identifier00883873871645
UDI-Public00883873871645
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Initial Date Received by Manufacturer 06/20/2018
Supplement Date Received by Manufacturer09/18/2018
Initial Report FDA Received Date07/09/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-