MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following physical damage on the insulin pump was noted: cracked battery tube thread, broken belt clip slot, broken reservoir tube lip, cracked reservoir tube, cracked case reservoir tube window corners, cracked reservoir tube window, cracked case near display window corners, missing end cap sticker, detached end cap and minor scratches on display window noted.
 
Event or Problem Description
The customer reported via phone call that their insulin pump was damaged.Customer stated the bottom plastic panel on the motor where the piston is popped out while rewinding.The customer¿s blood glucose level was 164 mg/dl at the time of the incident.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7672202
Report Number2032227-2018-08031
Device Sequence Number19175075
Product Code OYC
UDI-Device Identifier00643169202313
UDI-Public(01)00643169202313
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAH
Device Catalogue NumberMMT-523NAH
Device Lot NumberA000305621
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date07/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR
Patient Weight138
-
-