Model Number FAF07050
Medical Device Problem Codes Break (1069); Positioning Failure (1158); Retraction Problem (1536); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records were provided to the manufacturer.Images were provided.The lot number for the device was provided.The device history records and image are currently under review.The device was returned to the manufacturer for evaluation.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the stent graft was allegedly met with some resistance during deployment attempt and the delivery catheter allegedly broke.It was further reported that the stent graft allegedly did not deploy.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a stent graft placement procedure, the stent graft was allegedly met with some resistance during deployment attempt and the delivery catheter allegedly broke.It was further reported that the stent graft allegedly did not deploy.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was returned for evaluation.The device was visually evaluated and noted the outer catheter had a circumferential break.The inner lumen was also noted to be damaged/crimped.The device was unable to be functional tested.Therefore, the investigation is confirmed for break and material deformation as the inner lumen was crimped and the outer catheter was broke.The investigation is also inconclusive for partial deployment as the device was returned with the stent outside of the system.The investigation is also inconclusive for retraction issue as the event was unable to be reproduced.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAIR ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7672245
Report Number2020394-2018-01162
Device Sequence Number751874
Product Code PFV
UDI-Device Identifier04049519008790
UDI-Public(01)04049519008790
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P060002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFAF07050
Device Catalogue NumberFAF07050
Device Lot NumberANAR1097
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date07/09/2018
Supplement Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight70
-
-