DEPUY IRELAND - 9616671 ATTUNE CONV SHIM SZ7 20MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500945
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: examination of the returned device confirms the reported event of trial damage.The observed cracking is initiating around the metal insert of the trial.A pra (b)(4) (preliminary risk assessment) meeting was held to discuss a product escalation out of australia concerning the attune shims.The team has determined that there is no additional patient risk associated with this issue.Although cracks are occurring, this failure mode does not lead to a patient harm of infection.The rate of infection for attune is within the state rate in the risk management report and is statistically similar to the rate of infection for the total knee class.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the performed investigation and pra (b)(4) determination of no additional patient risk, corrective action is not indicated.Continue to monitor via sep-419.
 
Event or Problem Description
It was reported that instruments were broken.Didn't impact any case.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary examination of the returned device confirms the reported event of trial damage.The observed cracking is initiating around the metal insert of the trial.A pra(b)(4).(preliminary risk assessment) meeting was held to discuss a product escalation out of australia concerning the attune shims.The team has determined that there is no additional patient risk associated with this issue.Although cracks are occurring, this failure mode does not lead to a patient harm of infection.The rate of infection for attune is within the state rate in the risk management report and is statistically similar to the rate of infection for the total knee class.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the performed investigation and pra(b)(4) determination of no additional patient risk, corrective action is not indicated.Continue to monitor via (b)(4).If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV SHIM SZ7 20MM
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key7672973
Report Number1818910-2018-64026
Device Sequence Number11131474
Product Code HWT
UDI-Device Identifier10603295135227
UDI-Public10603295135227
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500945
Device Lot NumberBFA0R4C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Initial Date Received by Manufacturer 06/14/2018
Supplement Date Received by Manufacturer06/14/2018
07/11/2018
Initial Report FDA Received Date07/10/2018
Supplement Report FDA Received Date07/10/2018
08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-