MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD MONOSOF; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD MONOSOF; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Model Number SN-5666G
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the suture broke in half while tying a knot.They opened a new suture to complete the case.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, during finger amputation the suture broke in half while tying a knot, while wrapping around needle driver to tie.They opened a new suture to complete the case.The incident occurred at the end of the procedure.The case was finished on last stitch.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted one suture and one needle.The suture was returned broken.Unfortunately, as no sealed samples were returned, a tensile strength test could not be performed.A review of the device history record indicates this product was released meeting all medtronic quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOSOF
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7673311
Report Number9612501-2018-01267
Device Sequence Number13683022
Product Code GAR
UDI-Device Identifier10884521078765
UDI-Public10884521078765
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K900531
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberSN-5666G
Device Catalogue NumberSN-5666G
Device Lot NumberD7K0050X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2018
Initial Date Received by Manufacturer 06/25/2018
Supplement Date Received by Manufacturer08/08/2018
01/04/2019
Initial Report FDA Received Date07/10/2018
Supplement Report FDA Received Date10/03/2018
01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-