Catalog Number CXI-2.6-18-150-P-NS-ANG
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, in an endovascular revascularization of a chronic occlusion in the left anterior tibial artery, a cxi support catheter separated while the physician was using the "safari" technique.The separated length was approximately 3-4 inches long, and was successfully removed.Additional information regarding event details, patient anatomy and outcome has been requested, but is not available at this time.
 
Additional Manufacturer Narrative
Correction: additional information regarding the reported incident was received on the 3500 a medwatch form.When trying to dilate a vessel in the leg the catheter broke inside, about a 3-4 inch piece.It was able to be retrieved.It is a cook medical product.The package and the device were saved.The intended procedure: endovascular revascularization of chronic occlusion of the let anterior tibial artery utilizing safari technique with anterograde access and balloon angioplasty of the anterior tibial artery up to 2.5mm proximally and 2mm distally.Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, drawing, and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows one nonconforming event which could contribute to this failure mode; however, the affected unit was scrapped and not replaced.There was one other reported complaint for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CXI SUPPORT CATHETER
Common Device Name
KRA CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7673549
Report Number1820334-2018-02035
Device Sequence Number13805277
Product Code KRA
UDI-Device Identifier00827002500037
UDI-Public(01)00827002500037(17)200410(10)7815407
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K072724
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2020
Device Catalogue NumberCXI-2.6-18-150-P-NS-ANG
Device Lot Number7815407
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/25/2018
Supplement Date Received by Manufacturer09/11/2018
Initial Report FDA Received Date07/10/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age87 YR
Patient Weight63
-
-