| Model Number |
SA-85 |
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/12/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A heli-fx endoanchoring system was used as an accessory device with a non-mdt cuff for the endovascular treatment of a type ia endoleak of a non-mdt stent graft on.It was reported that four endoanchors were successfully implanted from the aortic cuff to the aortic neck.Aneurysm max diameter was 66mm and neck length was 12mm with a proximal neck angulation of 42 degrees.It was reported approximately 1 month post index procedure, a follow up ct scan showed three endoanchors no longer adequately penetrated the aortic wall.One endoanchor was in the dilated aorta and two were in the sac.No action was taken as a result of this event and the patient will be monitored by their physician.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|