Model Number H7493808015250
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon in two pieces.The balloon was loosely folded.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 56.2cm from the hub.The fracture faces were oval as if kinked prior to separation.There are numerous hypotube and shaft kinks.The tip is damaged.There are numerous hypotube kinks.Inspection of the remainder of the device presented no other damage or irregularities.There was no evidence of any damage or irregularities contributing to the reported crossing difficulty.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2018.It was reported that crossing difficulties were encountered.The target lesion was located in a coronary artery.A 2.5mm x 15mm quantum maverick balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7674012
Report Number2134265-2018-05937
Device Sequence Number11108582
Product Code LOX
UDI-Device Identifier08714729392767
UDI-Public08714729392767
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2020
Device Model NumberH7493808015250
Device Catalogue Number38080-1525
Device Lot Number0021485912
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2018
Initial Report FDA Received Date07/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-