| Catalog Number |
254500999 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the instruments were broken.Didn't impact any case.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirms the reported event.The larger balseal is damaged and still intact.Hhe (health hazard evaluation)/qrb (quality review board) (b)(4) recommended a device correction which was initiated on (b)(4) 2015.(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor per post market surveillance (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|