| Catalog Number |
2030-6525-1 |
| Medical Device Problem Codes |
Break (1069); Crack (1135); Fracture (1260)
|
| Health Effect - Clinical Codes |
Injury (2348); Osteolysis (2377)
|
| Date of Event |
10/14/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event regarding fractured screw was reported.The event was confirmed through the medical review.Method & results: product evaluation and results: not performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant.Concluded: "discussion: ¿2½-years post arthroplasty patient had a fall after which anterior hip and groin pain persisted.X-rays were described as normal.¿ [¿] a bone scan showed no obvious pathology, no evidence for loosening or infection.Cobalt and chrome levels were normal for an arthroplasty patient with 3.1-ug/l and <1.0-ug/l respectively.¿ ¿procedure-related factors.Suboptimal cup position regarding inclination and anteversion.Patient-related factors: patient trauma as secondary trigger event; high activity level as possible secondary factor but not confirmed.Device-related factors: none as also supported by explant photographic findings.No formal mar available because devices were discarded after revision surgery.Diagnosis: suboptimal cup position regarding inclination and anteversion has contributed to overload in the cup shell contributing to loss of fixation in the cup shell and consequent overload on the supplemental cup fixation screws.Osteolysis developed around the screw area causing osteolytic granuloma formation as soft tissue mass in the pelvis.During the first revision the problem was not recognized with only bearing exchange and thus the condition further deteriorated until catastrophic cup fixation failure with screw fracture only 9-months later requiring a second revision.¿ product history review: a device history review confirmed all devices accepted into finished goods conformed to specification.Complaint history review: no other events were reported for the lot indicated.Conclusions: the investigation concluded that loosening was caused by the shell position.The consulting clinician indicated in the medical review, "suboptimal cup position regarding inclination and anteversion has contributed to overload in the cup shell contributing to loss of fixation in the cup shell and consequent overload on the supplemental cup fixation screws.Osteolysis developed around the screw area causing osteolytic granuloma formation as soft tissue mass in the pelvis." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that surgeon revised patient's right hip due to a loose cup.Surgeon revised cup, liner head and sleeve.Postoperative diagnosis: patient underwent revision of primary right tha (b)(6) 2015 for metallosis.Patient underwent subsequent revision of acetabular component (b)(6) 2015.Update as per medical review:."the overload condition could continue its detrimental effects and caused persisting pain problems immediately after the first revision and continuing until catastrophic cup failure with loosening, screw fracture and severe bone loss requiring an extensive second revision surgery." suboptimal cup position regarding inclination and anteversion has contributed to overload in the cup shell contributing to loss of fixation in the cup shell and consequent overload on the supplemental cup fixation screws.Osteolysis developed around the screw area causing osteolytic granuloma formation as soft tissue mass in the pelvis.During the first revision the problem was not recognized with only bearing exchange and thus the condition further deteriorated until catastrophic cup fixation failure with screw fracture only 9-months later requiring a second revision.
|
| |
|
Search Alerts/Recalls
|