Catalog Number ATG120164
Medical Device Problem Codes Break (1069); Retraction Problem (1536); Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac vein and inferior vena cava, the pta balloon allegedly ruptured on the first inflation attempt.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.A visual inspection found a partial break at the proximal balloon weld.However, the balloon was not detached and no issues were noted to the distal balloon weld.Therefore, the investigation is confirmed for a break and is unconfirmed for the reported balloon detachment.The balloon was inflated and a pinhole rupture was noted on the proximal cone of the balloon.Therefore, the investigation is confirmed for the reported rupture.However, the investigation is inconclusive for the reported retraction issues as the balloon was unable to be refolded to it's original configuration due to the balloon damage.It is possible that the identified rupture could have contributed to the reported retraction issues.It is also possible that retraction issue could lead to the identified proximal weld break.However, the definitive root cause for the reported or identified issues could not be identified based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac vein and inferior vena cava, the pta balloon allegedly ruptured on the first inflation attempt.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7675001
Report Number2020394-2018-01165
Device Sequence Number16486376
Product Code LIT
UDI-Device Identifier00801741060823
UDI-Public(01)00801741060823
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K122984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATG120164
Device Lot Number93RC0043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer10/03/2018
Initial Report FDA Received Date07/10/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-