AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013465-100
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspection was performed on the returned device.The leak was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation determined the reported leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the non-tortuous, non-calcified, 95% stenosed, mid, left, superficial femoral artery (sfa) with a 3.00 x 100 mm armada 18 percutaneous transluminal angioplasty (pta) catheter.The pta catheter was advanced to the lesion and inflated.While the balloon was inflated, contrast was seen leaking from the strain relief at the hub.As they were able to sufficiently dilate the lesion and complete the procedure, a replacement pta catheter was not needed.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7675372
Report Number2024168-2018-05213
Device Sequence Number16532643
Product Code LIT
UDI-Device Identifier08717648207891
UDI-Public08717648207891
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number1013465-100
Device Lot Number6071441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-