| Model Number |
35102 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the adhesive on the male external catheters was too strong, making them difficult to remove even when rolling off and using a warm, moist compress.It was reported that the patient allegedly experienced pain while trying to remove the male external catheter.There was no medical intervention reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 1 unopened sample in the original packaging.No obvious defects were noted.An adhesive peel test was performed, the samples were cut ( 0.7100in) to the required width (0.70" +/- 0.05 ").And the sample was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication: the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply: 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove: gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
|
| |
|
Event or Problem Description
|
|
It was reported that the adhesive on the male external catheters was too strong, making them difficult to remove even when rolling off and using a warm, moist compress.It was reported that the patient allegedly experienced pain while trying to remove the male external catheter.There was no medical intervention reported.
|
| |
|