| Catalog Number |
ATG120164 |
| Medical Device Problem Codes |
Retraction Problem (1536); Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the iliac vein and inferior vena cava, the pta balloon allegedly ruptured on the first inflation attempt.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.Visual inspection was unable to identify anomalies or failures.The device was functionally tested and passed without issue.Therefore, the investigation is unconfirmed for retraction issue, material rupture and detachment.The definitive root cause for the reported issues could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the iliac vein and inferior vena cava, the pta balloon allegedly ruptured on the first inflation attempt.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|