• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2060-020
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Based on the information provided, the reported difficulties appear to be due to operational circumstances of the procedure.It is likely that the balloon rupture is the result of interaction with the lesion or associated devices during use.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the 6.0 x 20 mm armada 35 balloon ruptured during inflation.No adverse patient effect or clinically significant delay in the procedure was reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7675744
Report Number2024168-2018-05221
Device Sequence Number11151406
Product Code LIT
UDI-Device Identifier08717648155253
UDI-Public08717648155253
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue NumberB2060-020
Device Lot Number70914G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-