• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS STENT SYSTEM 3.0 X 21MM WITHOUT TIP; STENT, INTRACRANIAL NEUROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS STENT SYSTEM 3.0 X 21MM WITHOUT TIP; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Catalog Number M003SZAS30210
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Neurological Deficit/Dysfunction (1982); No Known Impact Or Consequence To Patient (2692)
Date of Event 05/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This is the first of 2 reports.Subject device remains implanted.
 
Event or Problem Description
It was reported that a patient underwent successful stent-assisted (y-stenting) coil embolization of an unruptured aneurysm located on the basilar apex.Approximately 8 month-post-procedure, re- intervention to target aneurysm was undertaken.The event was resolved with no residual effects.According to the clinical events committee (cec), the re-treatment was related to the procedure, related to the atlas stents (subject devices), and not related to the coil mass.Per site, the reason for the re-intervention was a large, expanding basilar aneurysm with compression on the brainstem causing symptoms.Aneurysm continued to expand despite atlas stenting and coiling.
 
Additional Manufacturer Narrative
Outcomes attributed to ae: added "other serious (important medical events)".Executive summary: based on the additional information received executive summary is updated to include, ¿per site, the reason for the re-intervention was a large, expanding basilar aneurysm with compression on the brainstem causing symptoms.¿ (b)(4).
 
Event or Problem Description
It was reported that a patient underwent successful stent-assisted (y-stenting) coil embolization of an unruptured aneurysm located on the basilar apex.Approximately 8 month-post-procedure, re- intervention to target aneurysm was undertaken.The event was resolved with no residual effects.According to the clinical events committee (cec), the re-treatment was related to the procedure, related to the atlas stents (subject devices), and not related to the coil mass.Per site, the reason for the re-intervention was a large, expanding basilar aneurysm with compression on the brainstem causing symptoms.
 
Additional Manufacturer Narrative
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device is not available; therefore, functional testing as well as physical analysis cannot be performed.In the case of this complaint it was reported that the patient had to go re-treatment of aneurysm 8 months after the stenting and coiling procedure due to aneurysm still expanding.However, recanalization requiring retreatment is a known risks associated with endovascular procedures and are noted as such in the device directions for use (dfu).Therefore, an assignable cause of anticipated procedural complications has been assigned to this event.Update: additional information received mentioned that the symptoms were "right sided weakness".
 
Event or Problem Description
It was reported that a patient underwent successful stent-assisted (y-stenting) coil embolization of an unruptured aneurysm located on the basilar apex.Approximately 8 month-post-procedure, re- intervention to target aneurysm was undertaken.The event was resolved with no residual effects.According to the clinical events committee (cec), the re-treatment was related to the procedure, related to the atlas stents (subject devices), and not related to the coil mass.Per site, the reason for the re-intervention was a large, expanding basilar aneurysm with compression on the brainstem causing symptoms of right sided weakness.It was reported that a patient underwent successful stent-assisted (y-stenting) coil embolization of an unruptured aneurysm located on the basilar apex.Approximately 8 month-post-procedure, re- intervention to target aneurysm was undertaken.The event was resolved with no residual effects.According to the clinical events committee (cec), the re-treatment was related to the procedure, related to the atlas stents (subject devices), and not related to the coil mass.Per site, the reason for the re-intervention was a large, expanding basilar aneurysm with compression on the brainstem causing symptoms of right sided weakness.After retreatment, at 8+ month post-procedure, the right sided weakness was ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS STENT SYSTEM 3.0 X 21MM WITHOUT TIP
Common Device Name
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7676264
Report Number3008881809-2018-00299
Device Sequence Number1554126
Product Code NJE
UDI-Device Identifier07613252656159
UDI-Public07613252656159
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
H020002/S046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Catalogue NumberM003SZAS30210
Device Lot Number19218065
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer06/18/2018
08/06/2018
Initial Report FDA Received Date07/10/2018
Supplement Report FDA Received Date07/17/2018
08/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age54 YR
-
-