| Catalog Number |
OL-XA25455M |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D6: implanted date: device was not implanted d7: explanted date: device was not explanted the status of the device is unknown at this time.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code 11 has been referenced in the conclusions section of h6.A review of the device history record and the shipping inspection records of the involved product code/lot# combination confirmed there was no indication of production-related problem or of discrepancy in the inspection result.A search of the complaint file did not find any other report with the involved product code/lot# combination.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.Exemption number e2015022.- attachment: [mw5077670.Pdf].
|
| |
|
Event or Problem Description
|
|
The user facility reported a fracture on the involved device through an fda mw5077670.The patient underwent an ercp (endoscopic retrograde cholangio-pancreatography) for biliary stent replacement.During the procedure the guide wires tangled in the biliary duct and were unable to be removed.The physician elected to cut off the wires and leave them in place.The needle knife used to trim the wires broke and the lever was retained in the patient.The procedure outcome is unknown.The final patient impact is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.With no device return the exact cause of the reported event cannot be definitively determined based on the available information.
|
| |
|
Search Alerts/Recalls
|