MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLANAV OI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLANAV OI Back to Search Results
Model Number M004R9620K20
Medical Device Problem Codes Deformation Due to Compressive Stress (2889); Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: visual inspection reveled that the distal end is severely twisted, especially immediately distal from the butt bond.The twist has clockwise rotation as viewed from the end of the distal tip.Body fluid ingress through damaged seals at the butt bond.Body fluid also seeped underneath both ring 1 seals.Dried body fluids were found on the handle, main body tubing, and on the distal tip.The luer fitting was cracked.The tension control knob functioned properly on both lock and unlock positions.The steering knob had very high resistance and could not be turned in either direction.Also, the distal end could not be curved in either direction, so curve tests could not be performed.Continuity checks revealed no electrical shorts or opens.-ray found spiraled wires at the distal end and collapsed guide coil at the weld to the center support.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is user / use error as there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user.The intellanav oi dfu ((b)(4)) states: "warnings: ¿ at no time should an intellanav¿ oi catheter be advanced or withdrawn when resistance is felt, without determining the cause.Valve damage, vascular and/or cardiac perforation is a risk with any intracardiac catheter.¿ catheter entrapment within the heart or blood vessels is a possible complication of cardiac ablation procedures.The potential for catheter entrapment may be increased when the catheter is over-torqued and/or positioned in the chordae tendineae.The occurrence of this complication may necessitate surgical intervention and/or repair of injured tissue and/or valve damage.¿ excessive curves or kinking of the catheter may damage internal wires and components, including the cooling lumen.This damage may affect steering performance and may cause patient injury.¿ manual bending and/or twisting of the distal curve can damage the steering mechanism and cooling lumens and may cause catheter failure and patient injury." and "precautions: ¿ the intellanav oi catheter is highly torquable.Avoid over-torquing.Over-rotating the handle and catheter shaft may cause damage to the distal tip or catheter assembly.Do not rotate the handle and catheter shaft more than one and one-half times the full rotation (540°).If the desired catheter tip position is not achieved, adjust the catheter¿s curve to disengage the catheter tip from the heart wall before resuming rotation of the handle and catheter shaft".(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 20june2018.A intellanav oi was selected for a 3d complex electrophysiology procedure.It was reported that catheter kinking and navigation failure occurred.No patient complications were reported.However analysis reveled seal break and butt bond failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV OI
Common Device Name
INTELLANAV OI
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
MDR Report Key7677559
Report Number2134265-2018-06067
Device Sequence Number1553693
Product Code OAE
UDI-Device Identifier08714729937449
UDI-Public08714729937449
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P150005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2020
Device Model NumberM004R9620K20
Device Catalogue NumberR9620K2
Device Lot Number21332456
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight68
-
-