• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.7MM CABLE WITH CRIMP 750MM-STERILE; CERCLAGE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 1.7MM CABLE WITH CRIMP 750MM-STERILE; CERCLAGE FIXATION Back to Search Results
Model Number 298.801.01S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information not available for reporting.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that three (3) separate cable tensioners have a cable stuck in the devices.Issue was discovered after an unknown procedure on (b)(6) 2018 and after the devices were cleaned.No patient harm was reported.This report is for one (1) cable with crimp.This is report 4 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.One unknown cable was received stuck inside a returned cable tensioner (product code: 391.201).The cable ends we beginning to fray, as the cable appeared to have been intentionally cut after the tensioner jammed and would not release the cable.Upon investigation at the supplier, the cable was able to be removed during a functional test of the tensioner.No issues were identified with the removed cable that would have contributed to the complaint condition.Additionally, the product code of the returned device was unable to be confirmed as not product marking was on the device.As no defect was identified with the cable following a visual inspection and functional test of the associated tensioner, no further investigation is required.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.7MM CABLE WITH CRIMP 750MM-STERILE
Common Device Name
CERCLAGE FIXATION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7677575
Report Number2939274-2018-52803
Device Sequence Number1370385
Product Code JDQ
UDI-Device Identifier10886982182500
UDI-Public(01)10886982182500
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K992616
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number298.801.01S
Device Catalogue Number298.801.01S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Initial Date Received by Manufacturer 06/12/2018
Supplement Date Received by Manufacturer09/11/2018
Initial Report FDA Received Date07/11/2018
Supplement Report FDA Received Date09/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-