Model Number S7
Device Problems Intermittent Loss of Power (4016); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Patient Involvement (2645)
Event Date 06/15/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information that, while outside of a procedure, the monitor of the navigation system would intermittently not complete start up.It was noted that the touch screen was unresponsive to prompts from the user when the monitor would start up.An audible beep was noted as well.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that replacing monitor resolved the issue.The system then passed the system checkout and was found to be fully functional.The monitor has not been received by the manufacturer for evaluation.
 
Manufacturer Narrative
The surgeon monitor was returned to the manufacture for evaluation.After functional testing and visual/physical examination the reported issue was not confirmed.The returned monitor displayed a good image with no anomalies when connected to a known good system.The touch function was found to be normal as well.No problem found.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7677620
MDR Text Key113751791
Report Number1723170-2018-03306
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450968
UDI-Public00613994450968
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2018
Initial Date FDA Received07/11/2018
Supplement Dates Manufacturer Received07/17/2018
07/29/2019
Supplement Dates FDA Received08/14/2018
08/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse