If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and inspected.Visual observation reveals the thumb screw was broken and the femoral rod is stuck inside the guide frame, confirming this complaint.It is possible that the screw was excessively tightened during procedure causing it to snap.Other than this possibility, we cannot discern a root cause for this failure.Although age of the device is unknown, it appears to be old and is in worn condition.This failure can be attributed to fair wear and tear.Furthermore, no lot number was supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|