The production device history record (dhr) for this iabp unit is not required to be reviewed per sop 01-061 since the device manufacture date is greater than one year from the event date.A getinge technician support (ts) replaced the following parts: the defective power supply; the main board due to broken jp15 connector; the broken charge led panel cable: the broken blood back tubing assembly.The ts also replaced the, fiber optic assembly due to "ap optical sensing module failure" as old module was not compatible with software upgraded datasettes.The 2500 hr pm kit was also installed & safety disk replaced.The solenoid driver board, gasket retro fit kit & software update field actions were also completed.(b)(4).
|