Model Number G29981
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pma/510(k) number: pre-amendment.(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during an angioplasty of the superficial femoral artery, the hub detached from the flexor ansel guiding sheath.Contralateral access was gained through the common femoral artery in the groin.There was no difficulty on insertion.As the physician was introducing an angioplasty balloon through the complaint device, the complaint device advanced inside the artery and the hub detached.The procedure was aborted in order to perform a surgical cut-down of the artery and manually remove the flexor ansel guiding sheath.No information was available regarding vessel tortuosity or calcification.No information has been reported regarding completion of the angioplasty procedure.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation a review of the dimensional verification, complaint history, device history record, documentation, instructions for use (ifu), manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the absence of a flare on the proximal end.Exposed coiling was also noted.The sheath liner was visible on the proximal end of the exposed coil.Measured from the proximal end, a kink was noted at a distance of 1.3 cm.The tubing was also observed to be flattened and compressed from a distance of 24.5 cm to 26 cm.The cause of the compression could not be identified.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
This mdr is being submitted as having information not previously reported.Additional complaint investigation and record remediation was not performed.Blank fields on this form indicate the information is unknown or unavailable.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7678323
Report Number1820334-2018-01986
Device Sequence Number19220897
Product Code DYB
UDI-Device Identifier00827002299818
UDI-Public(01)00827002299818(17)210228(10)8630607
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberG29981
Device Catalogue NumberKCFW-5.0-18/38-45-RB-ANL1-HC
Device Lot Number8630607
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/21/2018
Supplement Date Received by Manufacturer08/28/2018
03/19/2020
Initial Report FDA Received Date07/11/2018
Supplement Report FDA Received Date09/11/2018
03/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-