The device was not returned for analysis.The analysis is therefore based on the inspection of the quality documents associated with the manufacture of this particular device as well as the returned data.The returned data have been analyzed.The analysis revealed that the clinical observation resulted from a battery voltage lower than expected.However, the current consumption was found to be normal.Based on the information available for analysis, the root cause of this observation could not be determined and requires a device analysis.It cannot be excluded that this occurrence results from a component damage leading to a quicker discharge of the battery.A prediction of the remaining service time is thus not feasible.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process which might be related to the clinical observation.The final acceptance test proved the device functions to be as specified.In summary, the device was not returned for analysis.The review of the quality documents confirmed a regular device manufacturing.The analysis of the returned device data revealed a battery voltage lower than expected, leading to the clinical observation.For further insights regarding the root cause, an analysis of the device itself would be required.Should the device become available for analysis, the investigation will be updated.
|
|
Upon receipt, the device interrogation revealed the eos battery status, detected on (b)(6) 2018 at 01:39 am.The device was implanted for 62 months and 22 charging cycles were recorded in the devices memory.The current consumption of the icd was analyzed and found to be normal and as expected.However, an inconsistency between current consumption and battery voltage was noted.Therefore, the icd was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage confirmed a depleted battery.In a next step, the electronic module was attached to an external power supply and the eos status was removed with a technical programmer to check the functionality of the electronic module.There was no indication of a malfunction, particularly all therapy functions were available and worked as expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The visual inspection of the battery did not reveal any external signs of damage.The voltage measurement confirmed the battery depletion.At a next step, the battery was opened for destructive analysis.The inspection of the inner assembly identified damaged insulation, which led to an increased internal self-depletion within the battery and therefore contributed to the clinical observation.In conclusion, the device was implanted for 62 months.The therapeutic functionality of the device was tested with an attached external power supply and proved to be fully functional.The analysis revealed an increased internal self-depletion within the battery that contributed to the clinical observation.
|