• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICAN REPUBLIC INC. FILTERFLOW®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICAN REPUBLIC INC. FILTERFLOW®; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 473989
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A sample has not yet been received and the investigation is ongoing at this time.A follow-up report will be submitted when the results of the investigation become available.
 
Event or Problem Description
As reported by the user facility: during the application of chemotherapy, the head nurse observed leakage of taxol.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) unopened sample was returned for evaluation.Visual examination of the set noted zero (0) visual defects.The set was leakage tested, per specification, with passing results.Therefore, the reported defect was not confirmed.Review of the discrepancy management system (dsms) database performed for the reported lot number revealed no abnormalities or non-conformances noted during in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERFLOW®
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7678868
Report Number9614279-2018-00099
Device Sequence Number13665258
Product Code FPA
UDI-Device Identifier04045928016265
UDI-Public04045928016265
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K942988
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number473989
Device Lot Number0061608330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Initial Date Received by Manufacturer 07/03/2018
Supplement Date Received by Manufacturer07/03/2018
Initial Report FDA Received Date07/11/2018
Supplement Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-