• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US ANCHOR C SCREWDRIVER - STRAIGHT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US ANCHOR C SCREWDRIVER - STRAIGHT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 48328500
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/21/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the screwdriver tip broke during screw insertion.All broken fragments were retrieved and the procedure as successfully completed with an alternate instrument.There are no reports of adverse consequences to the patient.
 
Event or Problem Description
It was reported that the screwdriver tip broke during screw insertion.All broken fragments were retrieved and the procedure as successfully completed with an alternate instrument.There are no reports of adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported event was confirmed via functional inspection.The product was not inspected prior to surgery for wear or damage.Manufacturing files were reviewed and revealed to be over 3 years old with no anomalies found.The instrument was not excessively impacted.The patient had average bone quality.Anchor c surgical technique warns users that extensively used instruments are more susceptible to fracture and must be examined prior to surgery for wear or damage: "instruments which have experienced extensive use or extensive force are more susceptible to fracture depending on the operative precaution, number of procedures, disposal attention.Instruments must be examined for wear or damage prior to surgery." the plausible root cause of the reported event is most likely normal wear and tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANCHOR C SCREWDRIVER - STRAIGHT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7679312
Report Number0009617544-2018-00168
Device Sequence Number1316112
Product Code LXH
UDI-Device Identifier07613327267082
UDI-Public07613327267082
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48328500
Device Lot Number14E570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2018
Initial Date Received by Manufacturer 06/21/2018
Supplement Date Received by Manufacturer08/29/2018
Initial Report FDA Received Date07/11/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-