• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH NO 7. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH NO 7. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5537-G-709
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Injury (2348)
Date of Event 06/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding infection involving a triathlon insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned; medical records received and evaluation: no medical records were received for review with a clinical consultant ; device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot or sterile lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, pathology reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.A memo provided by a principal microbiologist concluded; due to the nature of the organism isolated - susceptible to various modes of sterilisation - and the sterilisation methodology employed by stryker, it is not possible that e.Coli could have originated from the implants.The following devices were also listed in this report: tibial baseplate; cat# unk; lot# unk; femoral component; cat# unk; lot# unk ; patella; cat# unk; lot# unk.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Not returned.
 
Event or Problem Description
It was reported that a stage 1 revision was performed on patient's right knee for infection (infection clinically confirmed, surgeon reported the infecting microorganism as e.Coli), after patient complaint of pain.A tibial baseplate, femoral component, patella (implanted (b)(6) 2018) and liner (implanted (b)(6) 2018) were all explanted and an antibiotic cement spacer placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NO 7. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7679427
Report Number0002249697-2018-02143
Device Sequence Number10969965
Product Code MBH
UDI-Device Identifier07613327027471
UDI-Public07613327027471
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Catalogue Number5537-G-709
Device Lot Number4L2P2M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2018
Initial Report FDA Received Date07/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age59 YR
Patient Weight108
-
-