| Model Number |
170043 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, while suturing in a laparoscopic exploratory laparotomy with hernia repair, the needle fell on the patient¿s cavity and was not retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The broken needle was returned lodged in the right jaw of the suturing device; the needle was broken at the suture swage with a mark on the cone of the needle.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the bent or broken needle may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|