| Model Number |
1000 |
| Medical Device Problem Code |
Failure to Run on Battery (1466)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's battery and verified the reported issue.Physio determined that cause of the reported issue was due to a failure of one of the battery cells.This caused the device to not be able to remain powered on and the device cannot charge and shock.The customer received a replacement battery.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had displayed a premature low-battery indicator.There was no patient use associated with the reported event.Upon further investigation, physio-control found that one cell in the battery has failed and the battery cannot perform correctly.As a result, the device cannot remain powered on to charge and shock.
|
| |
|