Model Number FUSION COMPACT |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unique device identifier (udi) is unavailable.Device manufacture date is unavailable.A medtronic representative went to the site to test the equipment.Testing revealed that replacing the emitter did not restore functionality.The representative then replaced the electromagnetic interface box and the communication cable the emitter for the navigation system was returned to the manufacturer for analysis.The unit was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|
Event Description
|
Medtronic received information that, while in a functional endoscopic sinus surgery (fess), the navigation system was experiencing difficulties tracking instruments.It was noted that the navigation system was restarted to temporarily restore functionality.After multiple restarts, the case could be completed as intended.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.
|
|
Manufacturer Narrative
|
The axiem was returned to the manufacturer for analysis.Analysis found that the issue could not be replicated.When connected to a test system with ent application registration, tracking and accuracy were normal on all ports.Connected and disconnected the tools from port and system remained normal.The axiem cable was returned to the manufacturer for analysis.Analysis found that the cable functioned normally without failure.Flexing the usb cables dies not indicate any intermittent opens.No fault found.
|
|
Search Alerts/Recalls
|