Blank fields on this form indicate the information has previously been reported, is unknown or is unavailable.Investigation evaluation: a review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, trends, quality control, and visual inspection of the returned device was conducted during the investigation.The complaint device was returned for evaluation.Visual inspection found no kinks or damage, however, a crack was found on the blue hub.During a leak test, the blue hub failed immediately.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually and functionally inspected by quality control and no notable gaps in production or processing controls were noted.The risk specification for this product includes the hub fracture failure mode and identifies that multiple risk controls are in place to mitigate the risk of this type of failure.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|